IMRT Multi-Slice Composite Plan Verification

This is a typical IMRT plan verification. The analyzed film in this example is a central axis section of a phantom, irradiated to a dose of 100cGy to isocenter. This film is compared to its associated dose matrix slice on the same phantom plan, planned for the same dose. The dose matrix was imported in DICOM RT-Dose format from the treatment planning system.

 

Total multi-slice comparison analysis can be performed – in parallel to –multi-slice 3D planning

 

 

 

 

 

 

 

The original case is a plan for the irradiation of the Cavernous Sinus. The dose distribution shape represented on the actual anatomy is very similar to that of the phantom’s

 

 

Adequately sized phantoms are recommended to avoid over-distorting the actual dose distribution

 

 

MLC QC and Calibration

MLC strips of 1 cm width are irradiated in sequence from left to right, with MLC banks traveling left to right.

 

An exposure is made every other 1 cm starting from –9.5cm to +11.5cm.

 

Every row of “little blue squares” should represent exposures of a leaf pair with an interleaf width of 1cm at various locations starting from –9.5cm to +11.5cm.

 

The small analysis line segment on the film – towards the mid-upper left – produces the profile to the right, which has precisely 10mm width at the 50% FWHM

 

 

The same analysis of a different interleaf pair separation – the yellow analysis line towards the upper right of the film – shows an inter leaf width of 7.42mm at FWHM. Indicating the necessity to undergo MLC re-calibration.

 

The dosimetric error between a correct width (top) and an incorrect one (bottom) is excessive in this case and can drastically affect dosimetric results of IMRT plans which employ small interleaf spacing.

 

The same analysis can be repeated for every leaf pair at positions spanning the area of interest. Calibration of leaf pair positioning can be made based on the 50% FWHM and peak to axis measurements.